Sign In | Join Free | My frbiz.com |
|
Categories | Gas Detector Sensor |
---|---|
Brand Name: | KACISE |
Model Number: | KTG111-C |
Certification: | CE |
Place of Origin: | CHINA |
MOQ: | 1PCS |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 1000 Piece/Pieces per Week negotiable |
Delivery Time: | 5-8 work days |
Packaging Details: | each unit has individual box and all boxes are packed in standard packages or customers requests available |
Measurement Range: | 0.00 - 2.00%/25.0% (Optional measurement ranges) |
Allowable Error: | ±2%FS |
Repeatability: | ±1%FS |
Resolution: | 0.0001 |
Response Time: | t90 ≤ 30s |
Zero Drift: | ≤±1%FS/7d |
Company Info. |
Xi'an Kacise Optronics Co.,Ltd. |
Verified Supplier |
View Contact Details |
Product List |
The KTG111-C oxygen in hydrogen analyzer is applied to water electrolysis for hydrogen and oxygen production
Product Applications
The product is widely used in on - line detection and analysis of hydrogen content in explosion - proof and non - explosion - proof scenarios during industrial production processes such as water electrolysis for hydrogen and oxygen production, fertilizers, petrochemicals, air separation, and biological fermentation.
Basic Principle
This instrument uses a fuel - cell - type electrochemical sensor, usually composed of silver and lead electrodes immersed in an electrolyte. The measured gas diffuses through a gas - permeable membrane to the electrodes, where a chemical reaction occurs, enabling the measurement of oxygen content.
Technical Parameters
Measurement Range: 0.00 - 2.00%/25.0% (Optional measurement ranges)
Allowable Error: ±2%FS
Repeatability: ±1%FS
Resolution: 0.01%
Response Time: t90 ≤ 30s
Zero Drift: ≤±1%FS/7d
Span Drift: ≤±1%FS/7d
Sample Gas Flow Rate: (300 ± 30) mL/min
Operating Power Supply: Transmitter 24VDC
Operating Environment:
Operating Temperature: - 15℃ - +45℃
Operating Humidity: ≤90%RH
Instrument Features
Intrinsic Safety Type
Overall Dimensions: 140mm (Width)×140mm (Height)×220mm (Depth)
Cutout Dimensions: 136mm (Width)×136mm (Height)
Parameters of External Safety Barrier: Vo ≤ 28VDC, Io ≤ 93mA, Cp ≤ 0.02uF, Lp ≤ 1mH
Weight: 1.8kg
Flameproof Type
Dimensions: 130mm*136mm
Analog Output: 4 - 20mA
Digital Output: RS458 interface, modbus protocol.
Explosion - proof Mark: Exd IIC T6
Weight: 1.8kg
![]() |