Measurement range: 0.1mg/L~10000mg/L~%
Sample state can be measured: Solid, liquid, gas(corresponding sample injector is optional)
PMT high voltage range: DC400V~900V(according to the level of the measured concentration, set the required value through the operation system.)
Temperature control range: Room temperature to 1200℃
Temperature control accuracy: ±1℃
Repeatability error: 1.0mg/L<X≤10mg/L, Cv≤±10% 10mg/L<X Cv≤±5%
Configuration: Host (data acquisition, control system), cracking furnace, autosampler and computer (operating software included).
Host size: 550 ×560× 200 (mm)
host weight: 9kg
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

ASTM D5453 Ultraviolet Fluorescence UV Sulfur Analyzer AC220V±10% 50HZ

SH0689 UV fluorescence sulfur meter

SH0689 Ultraviolet Fluorescence Sulfur Analyzer uses ultraviolet fluorescence method to determine the principle.

In line with ASTM D5453, SH/T 0689-2000 and other standards, suitable for the determination of total sulfur content in crude oil, distillate oil, liquefied petroleum gas, plastics, petrochemical products, food and other.

After the sample was introduced into the high temperature cracking furnace, the sample underwent cracking oxidation reaction.

At a high temperature of about 1000℃, the samples were completely gasified and oxidized, and the sulfur compounds in them were quantitatively converted to sulfur dioxide.

Ultraviolet Fluorescence ASTM D5453 Sulfur Analyzer Biodiesel Sulfur Content Tester

Reaction gas carried by carrier gas, through the membrane dryer to remove the moisture, into the reaction chamber.

Sulfur dioxide is irradiated by ultraviolet light at certain wavelengths and absorbs this radiation to divert some electrons into high-energy orbits.

Once the electrons have returned to their original orbitals, the excess energy is released as light, detected by a photomultiplier tube at a specific wavelength, emitting fluorescence that is completely specific to the sulfur and proportional to the amount of sulfur in the original sample.

After being amplified by micro-current amplifier and processed by computer, it can be converted into electrical signal proportional to light intensity, and the sulfur content of the corresponding sample can be calculated by measuring its size.

Scope of application: It is suitable for the determination of total sulfur content in gasoline, diesel oil, lubricating oil, crude oil, distillate oil, liquefied petroleum gas, gas, natural gas, plastics, petrochemical products, food, coal and so on.

Compliance with standards: ASTM D5453, SH/T 0689-2000 and other standards.

(It meets the testing standard requirements of national five national six gasoline diesel)

ASTM D5453 Ultraviolet Fluorescence UV Sulfur Analyzer

Lab Testing Machine Widely Use for Sulfur Content in Diesel by UV Fluorescence (ASTM D5453)

Sulfur Content in Diesel Oil Analyzer by UV Fluorescence

ASTM D5453 D4629 Total Low Sulfur Content in Oil Test Machine

ASTM D5453 Trace Sulfur Content Analyzer in Hydrocarbon Samples


Technical Parameters

Measurement range0.1mg/L~10000mg/L~%
Sample state can be measuredSolid, liquid, gas(corresponding sample injector is optional)
StandardASTM D5453 SH/T 0689-2000
PMT high voltage rangeDC400V~900V(according to the level of the measured concentration, set the required value through the operation system.)
Temperature control rangeRoom temperature to 1200℃
Temperature control accuracy±1℃
Repeatability error

1.0mg/L<X≤10mg/L, Cv≤±10%

10mg/L<X Cv≤±5%

ConfigurationHost (data acquisition, control system), cracking furnace, autosampler and computer (operating software included).
Host size: 550 ×560× 200 (mm) host weight : 9kg
Temperature control size: 550 ×560× 400 (mm) temperature control weight: 10kg

ASTM D5453 ASTM D7183 Low detection limit of UV fluorescence sulfur analyzer
Sulfur in Oil Analyzer by Ultraviolet Fluorescence ASTM D5453 Fluorescence Sulfur-in-Oil Analyzer Ultra Low Sulfur Analysis by ASTM D5453 UVF Method

The main features

1. computer control operation, can realize man-machine dialogue, data, peak shape can be automatically stored and printed.

2. The operating software can be connected with LIMIS system to realize automatic data transmission.

3. The instrument adopts a new type of quartz cracking tube, which can increase the injection volume and reduce the measurement error.

4. the cooling fan automatic switch, save the trouble of waiting for the sample shutdown, save time and effort.

5. The darkroom system adopts the latest foreign technology, and the key components adopt imported detection components, which improves the detection accuracy of the instrument

6. the instrument adopts imported membrane dryer to absorb moisture, saving the trouble of replacing magnesium perchlorate frequently, and reducing the problem of air leakage.

7. the use of computer control temperature, improve the instrument temperature control precision, and not easy to overtemperature.

8. the use of air cooling instead of water cooling, save water resources.

9. you can take the existing calibration curve with standard sample calibration for sample analysis.

Technical parameters

Determination range: 0.1 ~ 10000mg/L (PPM)

Lower detection limit: 0.1ppm (sulfur)

Repeatability error: 1.0mg/L < X≤10mg/L, CV ≤±10%

10 mg/L < X Cv plus or minus 5% or less

Air source requirements: high purity oxygen: more than 99.98%;

High purity argon: more than 99.98%

Temperature control range: room temperature ~ 1200℃

Temperature control accuracy: ±1℃

Working power supply: AC220V±10% 50Hz

Environmental requirements: temperature 15 ~ 40℃;

Humidity of 85% or less

Packing list

S/NNameUnitQuantityRemarks
1Hostset1
2Samplerset1
3computerset1With 232 serial port
4Quartz tubepiece1provided by the installer
5Elbowpiece1provided by the installer
6Pipe joint clamppiece1
7Quartz tube fixing clip (ceramic) ringpiece2
8Polytetrafluoroethylene pipemeter10
9Silicone rubber tube (inner diameter) Ф5meter0.2
10Ф3meter0.2
11Membrane dryerpiece1Imported from Japan
12power cordpiece2
13Power line of cracking furnacepiece1
14Communication cablepiece1
15Air line pressing cappiece2
16Silicone rubber padpiece20
17Sampling probe(10ul,25ul)piece各2
18standard samplebox4Installed with
19Fuse tube 20Apiece4
202Apiece4
21Warranty card certificatecopy1
22instruction manualcopy1

Approximate conversion formula

1%=10000PPm

1ppm=1mg/L

Quality ASTM D5453 Ultraviolet Fluorescence UV Sulfur Analyzer AC220V±10% 50HZ for sale
Send your message to this supplier
 
*From:
*To: Shandong Shengtai instrument co.,ltd
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0