Sign In | Join Free | My frbiz.com |
|
Categories | Oil Analysis Testing Equipment |
---|---|
Brand Name: | Skyline |
Model Number: | SL-X100 |
Certification: | CE, Rosh, Calibration |
Place of Origin: | China |
MOQ: | 1 |
Price: | Negotiable |
Payment Terms: | T/T, L/C, D/A, D/P, Western Union, MoneyGram, In Cash |
Supply Ability: | 30 unit / month |
Delivery Time: | 7 working days |
Packaging Details: | plywood case |
Sulfur measurement range: | 0.0017% ~ 5%; 1%-80% (two options) |
Measurement time: | 60-180 seconds |
Single sample automatic measurement, measurement repetition times: | 1, 2, 3, 5, 10 arbitrary setting, the end of the measurement to give the average and standard deviation |
Repeatability (r): | < 0.4347 X0.6446 (X= the average of two repeated test results) |
Reproducibility (R): | < 1.9182 X0.6446 (X= the average of two single, independent test results) |
Sample size: | 5ml ~ 6ml |
Storage: | can store 10 calibration curves |
Working conditions: | Temperature :10 ~ 30℃; Relative humidity :≤85% (30℃) |
Power supply: | AC100-220V/50Hz; Rated power :50W |
Size and weight: | 460mm*260mm*260mm; 11kg |
Company Info. |
SKYLINE INSTRUMENTS CO.,LTD |
Verified Supplier |
View Contact Details |
Product List |
X-ray fluorescence sulfur analyzer
Product introduction
Energy dispersive X-ray fluorescence sulfur analyzer is a special analyzer based on energy dispersive X-ray fluorescence spectroscopy. The analyzer principle, design structure, instrument elements, analytical method operation flow, error and reproducibility all conform to GB/T 17040-2019 and ASTM D4294 for the determination of sulfur content in petroleum and petroleum products.
Main use
◆ Measuring sulfur content of crude oil, petroleum, heavy oil, diesel oil, kerosene, naphtha;
◆ Measuring the total sulfur content of coal chemical products, such as primary benzene;
◆ Measurement of sulfur content in other liquid or solid fine powder samples such as petroleum coke.
Instrument characteristics
◆ Intelligent electromechanical integrated microcomputer design,8 inch wide viewing Angle capacitive touch screen (1024*768) display
◆ Wide detection variety, wide detection range, fast analysis speed, less standard sample consumption;
◆ The instrument has automatic peak stabilization function and automatic diagnosis function, monitoring the working state and electrical parameters of the instrument;
◆ Disposable Mylar film sample cup to avoid cross contamination; Sample cup production is convenient and fast;
◆ Accurate positioning of the sample table, easy to place the sample and replace the anti-oil leakage film, and avoid the possibility of contamination of the detection system;
◆ Large storage capacity of instrument data, 4096 content analysis results and 8192 count measurement data, 16 instrument calibration results.
◆ Thermal printer automatically prints test results;
◆ The instrument automatically selects the working curve without user intervention;
◆ The user can choose the measurement result unit, ppm or (m/m)%;
◆ Safe radiation protection measures, no harm to individuals and groups of ionizing radiation.
Technical index
1. Sulfur measurement range: 0.0017% ~ 5%; 1%-80% (two options)
2. Measurement time: 60-180 seconds;
3. Single sample automatic measurement, measurement repetition times: 1, 2, 3, 5, 10 arbitrary setting, the end of the measurement to give the average and standard deviation;
4. Repeatability (r): < 0.4347 X0.6446 (X= the average of two repeated test results);
5. Reproducibility (R): < 1.9182 X0.6446 (X= the average of two single, independent test results);
6. Sample size :5ml ~ 6ml;
7. The instrument can store 10 calibration curves;
8. Working conditions: Temperature :10 ~ 30℃; Relative humidity :≤85% (30℃); Power supply :AC100-220V/50Hz; Rated power :50W;
9. Size and weight: 460mm*260mm*260mm; 11kg.
![]() |