Accessories Included | USB charger, power cable, and clips |
Automatic Buffer Recognition | Yes |
Backlight | Yes |
Calibration Intervals/Alerts/Reminder | Off, selectable from 2 hours to 7 days |
Certifications | CE, FCC, ISED, RCM, KC, ETL Verified: US DOE/ NRCan Energy
Efficiency, RoHS |
Data Export | USB connection to PC or USB storage device (limited to the storage
device capacity). |
Data Memory | 5000 data points |
Data Storage | Automatic in Press to Read Mode and Interval Mode. Manual in
Continuous Read Mode. |
Dimensions (H x W x D) | 63 mm x 97 mm x 220 mm |
Display | Up to 3 parameters at a time, dependent on HQ model |
Display Type | 536x336 Mono-TFT |
Enclosure Rating | IP67 (with battery compartment installed) |
GLP Features | Date; Time; Sample ID; Operator ID, Calibration |
Inputs | 1 |
Instrument | Portable |
ISE Direct Measurement Range | No |
ISE Electrode Calibration | No |
Kit? | Meter only |
Languages | English, Chinese - Simplified, Spanish, French, German, Italian,
Japanese, Turkish, Dutch, Portuguese (PT & BR), Korean, Danish,
Swedish, Polish, Norwegian, Hungarian, Greek, Finnish, Czech,
Romanian, Croatian, Bulgarian, Slovak, Lithuanian, Estonian,
Slovenian, Russian |
Lock Function | Continuous / Auto-stabilization ("press to read") / At Interval |
Measurement Method | Probe specific programmed method settings |
Model | HQ1110 pH/ORP/1 Channel |
mV Measurement at Stable Reading | Yes |
mV Measurement Range | Yes |
mV Resolution | 0.1 mV |
Operating Error Messages | Clear text error messages displayed |
Operating Humidity | 90% (non-condensing) |
Operating Interface | Soft Touch Keypad |
Operating Temperature | 0 - 60 °C |
ORP Electrode Calibration | Predefined ORP standards (including Zobell's solution) |
Parameter | pH/ORP |
pH Buffer Sets | Colour-coded: 4.01, 7.00, 10.01 pH; |
IUPAC: 1.679, 4.005, 7.000, 10.012, 12.45 |
DIN: 1.09, 4.65, 9.23 |
User-defined custom buffer sets |
pH Electrode calibration | 1 - 3 Calibration points |
Calibration summary data logged and displayed |
pH Measurement Range | 0 - 14 pH |
pH Resolution | Selectable: |
0.001/0.01/0.1 pH |
Power Supply | Rechargeable lithium-ion battery 18650 (internal) |
Class II, USB power adapter: 100 - 240 VAC, 50/60 Hz input; 5 VDC
at 2 A USB power adapter output (external) |
Probe Type | Intellical Standard Laboratory or Rugged Field |
Probes included | None |
Protection Class | IEC Class III (SELV (Separated/Safety Extra-Low Voltage) powered);
external |
Sensor A | None |
Sensor B | None |
Sensor C | None |
Storage Conditions | -20 - 60 °C, max. 90% relative humidity (non-condensing) |
Temperature Compensation | Automatic Temperature compensation for pH |
Temperature Measurement | °C or °F |
Temperature Resolution | 0.1 °C |
Type of measurement | Field Measurements |
Warranty | 12 months |
Weight | 519 g (Meter only) |
What's included? | HQ1110 Portable pH/ORP Meter, 1 channel, includes manuals, Li-Ion
batteries, USB charger with cable for USA. |