Sign In | Join Free | My frbiz.com |
|
Categories | Hospital Emergency Equipment |
---|---|
Brand Name: | SUCE |
Model Number: | OEM |
Place of Origin: | China |
Packaging Details: | Standard export package |
Supply Ability: | 3000 Unit/Month |
Price: | $6,800.00 - $9,000.00/Units |
MOQ: | 1pc |
Payment Terms: | T/T, Western Union,Payoneer |
Delivery Time: | about 10 days |
Certification: | CE,ISO13485 |
Type: | Body Fluids Processing Equipments |
Instrument classification: | Class II |
Warranty: | 2 years |
After-sale Service: | Online technical support |
Product name: | Hemodialysis machine |
monitor: | 15" touch-screen monitor |
Power: | 220 Volts AC, 50 Hz |
Material: | All wetted parts in stainless steel |
Pressure monitoring: | The following pressures are monitored |
Inlet pressure min: | min. 3 bar at 1000 l/h to 3000 l/h dynamic |
Inlet pressure max: | 6 bar static |
Storage temperature: | 5℃~40℃, at relative humidity<=80% |
Keyboard: | 4 button handling |
Port: | SHANGHAI |
CE approved!! Double pump multifunctional Hemodialysis machine/ DIALYSIS machine HM02
Applications:
*Multifunctional hemodialysis machine is used to clear the toxin in the blood. It is also used for the patients with poisoning, renal failure, diabetes, kidney disease and uremia.
Techinical Specification:
- Open system allowing the utilization of consumables from
different suppliers
- 15" touch-screen monitor, rotatable, display of icons, words for
quick and easy use
- Operating instructions shown on the high-resolution 15" TFT
monitor
- Volumetric ultrafiltration control and sequential ultrafiltration
- Integrated status indicator in the monitor
- Fully automatic self-test prior to any start of dialysis
automatic priming
- Heparin pump for syringes from 10 to 30 ml, programmable infusion
(no infusion, profiled infusion, auto-bolus or continuous infusion)
-Bicarbonate powder cartridge holder: for dialysis treatments with
cartridges for hygienic dialysis
-Blood leak and air bubble detector
-Continues on-lin Kt/V measurement without the need of any
disposable supplies
-Automated temperature and pressure measurement
- Six individually programmable and/or preset profiles for
Dialysate temperature
-Sodium
-Ultra-filtration
-Bicarbonate
-Dialysate flow
-Heparin rates
-Dialysis flow rate adjustment from 300 to 800 ml/min in individual
steps of 1 ml
-Pyrogene and endotoxin filter to produce ultrapure dialysis fluids
-One blood pump with adjustable settings
-Battery power supply in case of power breakdowns (15 minutes
d'autonomie)
-Automated blood pressure measurement (ABPM), non-invasive by
oscillometric method, various programmable cycles
-Biologic RR Comfort, automated blood pressure stabilisation based
on blood pressure related ultrafiltration control without the need
of any additional propriety disposable supplies Crit-Line interface
connection allows the connection of the Crit-Line device to the
Dialysis Machine. -Crit-Line is a fluid management access
monitoring tool. It measures the absolute hematocrit in percent,
assesses haemoglobin, measures continuous oxygen saturation and
vascular access recirculation Patient disc and card reader: this
option registers automatically the patient at the machine,
activates automatically the set-up of all patient individual
dialysis parameters and saves treatment data on the patient card
-Single-needle dialysis as SN click-clack (requires no specific
blood tubing)
-Pressure sensor (PBE) to identify coagulation in dialysers prior
to any transmembrane pressure (TMP) alarm
- Integrated timer on/off for automated disinfections/preparation
procedures
- Programmable cleaning and disinfection modes, for chemical, heat
or a combination of chemical/heat disinfection, programmable weekly
disinfection table, validated heat-disinfection for citric acid 50%
(bactericidal, fungicidal, tuberculocidal, virucidal, inactivation
incl. HBV, HCB, HIV)
- Automated standby function to safe permeate, concentrates and
energy in preparationmode
- Complies with EN 60601
- CE approved
- Power: 220 Volts AC, 50 Hz.
- Plug type: CEE 7/7
Optional:
- Same configuration as above but with double pump
Water Treatment System Two unit each Can Support Twenty
Heomodilaysis Machines
2.1 Per-filtration system including:
-Back flush filter and pressure-reducing valve
-Booster pump
-Sand Filter System and / or
-CIR Filter System for remove of ferric, manganese, sulphur in
clear drinking water
-Activated Carbon Filter System For rest dechlorination of drinking
water
-Chlorine Filter for treatment of chlorinated portable water
-Double softener system
-Media rail panel:
- Aluminium channel with dead zone free permeate connections.
- All connectors are heat-disinfectable and resistant for hot
water.
- Single place or double place connectors available for permeate,
drain water and concentrate.
- The panel includes a centralized drain column system – which is
covered by the same Aluminium cover as the main panel itself, it
prevent smell and noise.
- Inside the panel there is an isolation to keep the Pe-Xa loop
constantly heated up while a heat disinfection process is on-going.
- Upgrades of additional connectors and extensions are easily
possible because of the compact design and the flexible components.
- Ultrafiltration and Heat-disinfection
- Ultrafiltration guarantees a reliable retention of bacteria and
endotoxines. The heat-disinfection enfolded the whole permeate
ring-main as well as the dialysis connection pipe inclusive the
inlet sector.
Equipment / Specifications
Patented procedure of Ultrafiltration of permeate
High lifetime of the ultra-filter by cross-flow-operation
Water economy by concentrate recirculation of the ultra-filter to
the reverse osmosis
3-step permeate pressure valve for fully automated filling of the
tank at permeate oversupply
Micro controller processed
Integrated pressure pump
Compatible for reverse osmosis systems
Hygienic
- Complete V4A-tubing, inside surface refined with clamp connectors
- PVC-free assembly of the tubing
Performance data:
- Tank volume: 500 litres, usable water quantity ca. 480 litres
- Double walled stainless steel tank (1.4404) with Isolation
- Ultrafiltration membrane made of polyether sulphone
Retention:
-Bacteria retention > 5 log step (Factor 100.000)
-Endotoxin retention > 5 log step (Factor 100.000)
-Temperature at heat disinfection > 75°C
Mechanical connections:
- Permeate input, permeate reduction reverse osmosis, permeate
ring-main feed, permeate return, 3 metres pressure tube DN 20,
drain HT 50 (com. to ring feeding tube connector and ring return
connector)
Electrical data:
- 400 V, 3~/N, 50-60 Hz, max. 14 kW
- Heater 12 kW, pressure pump max. 2 kW
- Dimensions / Weight (WxHxD): 790 x 1950 x 1450 mm / ca. 350 / 850
kg (tank empty / full)
Optional:
Water Protector
- Temperature sensor for measuring of temperature at the beginning
of the ring
- Cutting unit for shut-off of the permeate supplying during the
chemical disinfection of the reverse osmosis without opening the
ring-ain.
Manufactured to MPG/MDD – System of group IIa
2.2 Central reverse osmosis unit
- All pipe work manufactured in stainless steel. Free of dead legs
at the conduit and connectors which grants permeate of highest
purity (purified water).
- Products graded in permeate capacity / extension class:
350 l/h, 700 l/h, 1050 l/h, 1400 l/h, 1750 l/h
- Permeate capacity each at 12°C water temperature
Upgradable by extension sets.
- For increased performances several systems can be connected in
parallel.
Water inlet:
- Inlet pressure min.: min. 3 bar at 1000 l/h to 3000 l/h dynamic
- Inlet pressure max.: 6 bar static (pressure reducer in front of
the system required)
Voltage:
- 230/400V 3-phase 50 Hz
- 230/400V 3-phase 60 Hz
- 115/200V 3-phase 60 Hz
- Power consumption electric: 3 kW
Used Guidelines, standards and regulations:
- Guideline of Council 93/42 EWG Medical Device Directive
- Medical electrical equipment – Part 1: General requirements for
safety (IEC 601-1:1988 + A1:1991 + A2:1995)
- Safety requirements for electrical equipment for measurement,
control, and laboratory use part 1: General requirements (IEC
61010-1:2001)
Temperature sensor:
- Monitored temperature at permeate for system protection.
Optional overflow valve:
-An additional overflow valve made of stainless steel provides
over-temperature protection of the ringmain and dialysis machine.
- Pressure monitoring: The following pressures are monitored:
- Permeate pressure electronic
- Inlet pressure electronic and by pressure gauge
- Pump pressure by pressure gauge
- Conductivity measurement: Plug-in conductivity sensors monitored
the following conductivities:
- Softened water
- Permeate
- Fine wire fuses: Protects the control and monitoring circuits
against possible overload.
- Short regeneration time (approx. 25 minutes) (DVGW approved).
- 20" Pre-filter System for removal of medium and smaller
particulate matter.
- Housing material: PP pressure resistant up to 8.6 bar
- Filter insert: 5 u m absolute cartridge
- Ball valve for soft water control and sampling.
- Frame and break tank manufactured in prickled stainless steel
(AISI 304).
Pressure Pumps
Type:
-Grundfoss CRi 3-25 (50 Hz) / CRi 3-31 (60 Hz)
-Grundfoss CRi 3-17 (50 Hz) / CRi 3-21 (50 Hz)
Material:
- All wetted parts in stainless steel
Type of pump:
- Vertical, normal suction pump.
- With maintenance-free sliding seals.
- Connection by hygienic clamp-connectors.
Membrane housing
-Material: stainless steel AISI 316L. Matt pickled outside and
inside. 4"membrane
-Membrane equipped with thin film polyamide membranes 4"
-DOW XLE-4040 – polyamide wounded – Extra Low Energy
-Retention rate of inorganic substances: > 98%
-Pipe work in stainless steel AISI 316L / AISI 316Ti. Matt pickled
outside and inside.
-Ball valve in stainless steel with position permeate to ring main
resp. permeate to drain for service requirements.
Main controller consisting of:
- Control board: Performs all control and monitoring functions.
Self-checking on start-up
- Text display: Provides menu, guidance, displays for alarms,
operating conditions and limit values.
- Keyboard: 4 button handling.
-Optional remote control controlling and monitoring from a central
control room.
![]() |