Categories | HPLC HbA1c Analyzer |
---|---|
Brand Name: | LABNOVATION |
Model Number: | LD-560 |
Certification: | IFCC, NGSP, CE |
Place of Origin: | Shenzhen, China |
MOQ: | 1 Unit |
Price: | USD7,000.00-12,000.00/Unit |
Supply Ability: | 100 Units/Day |
Delivery Time: | 15-20 working days |
Packaging Details: | Carton Size: 480mm×480mm×615mm, Gross Weight: 28kgs |
Principle: | HPLC (High Performance Liquid Chromatography) |
Test Speed: | 90s/test |
CV: | <2% |
Sample types: | Whole blood or Pre-diluent sample |
Test blood volume: | 5μL in whole blood mode; 150μL in pre-diluted mode |
Sample loading method: | Continuous automatic sample loading |
Sample capacity: | 20 samples |
Reagent pack: | 600T |
Column test quantity: | 3000T |
Eluent Kind: | Eluent A, Eluent B,Eluent C (Variant Mode) |
Data storage: | 5000 |
Printer: | Built-in Thermal Printer |
Test range: | 3.0%~18.0% |
Operation screen: | 10" Full color LCD display |
Gross Weiht: | 28kg |
Dimensions: | 378(L) ×380(W)×510(H)mm |
Shelf Life: | 5 Years |
Warranty of Device: | 1 Year |
Warranty of Sampling Sensor: | 6 Month |
Company Info. |
Labnovation Technologies, Inc. |
View Contact Details |
Product List |
Automatic HbA1c Analyzer High Performance HPLC Golden Method for Diabetes Diagnosis
Intended use
Specification
Model | LD-560 |
Methodology | High-Performance Liquid Chromatography (HPLC) |
Test Modes | Fast Mode, Variant Mode |
Test Range | 3% - 18% |
Precision (CV) | <2% |
Parameter | HbA1c, HbF, HbA1a, HbA1b, HbA0, HbA2(Optional) |
First Sample Result | 4 Mins |
Test Speed | Fast Mode: 90 Seconds/Test Variant Mode: 140 Seconds/Test |
Sample Type | Whole blood, Pre-diluted blood |
Sample Volume | Whole blood: 5μL Pre-diluted blood: 150μL |
Sample loading capacity | 20 |
Photometer | 420nm |
Chromatography Column Capacity | 3000T |
Filter Capacity | 600T |
Display | 10“ LCD touch screen |
Reagent Kit | Eluent A, Eluent B, Eluent C, Hemolysis, Chromatography Column |
Storage | 5000 Sample Results |
Connection interface | USB, LAN, LIS Compatible |
Printer | Thermal Printer |
System operation environment | Temperature 15-30℃ Humidity 30%-85% |
Power | AC 100-240V 50/60HZ 300VA |
Dimensions | 378mmx380mmx510mm |
Weight | 28kgs |
*HPLC, CV<2% - Accurate and Stable
*Full Automatic smart System
*Quality & Performance - IFCC & NGSP Certificate
*User Friendly
Main Parts
The analyzer consists of an auto sample injection module, a liquid
chromatography separation module, a colorimetric detection and
control module, a barcode scanning module, and a display printing
module.
Reagents and Consumables
Eluent A/Eluent B/Eluent C(Variant Mode) | Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin. |
Hemolysis Reagent | Used to dissolve red blood cells and release hemoglobin. |
Chromatography Column | Used for separation and elution of glycosylated hemoglobin and non glycated. |
Hemoglobin Filter | Used to initially filter the sample after hemolysis. |
Paper Rolls | Prints a report of the test results. |
Certificate
![]() |